Important:
Before you use your PharmaJet Stratis® Needle-Free Injection device, please read this User Guide and the Instructions For Use completely. Follow the directions carefully when using the Stratis® device.
Refer to Instructions for Use to ensure safe injections and to review risks.
Stratis® System Components
Work Flow
1. Prepare the Injector
Place Injector nose first into the Reset Station. Push Injector forward to fit into place. Make sure the pin is in line with Injector.
STAND and push FIRMLY on the lid of the Reset Station.
Continue to push down firmly against the resistance until lid closes completely. Lid will lift indicating Injector is charged.
2. Fill Syringe
Remove blue protective cap from Syringe.
Place Syringe into top of Vial/Filling Adapter. Push down and twist to lock Syringe into place.
Insert spike downward into vial with a twisting motion.
Invert Adapter/Syringe assembly, pull down on Syringe plunger to draw fluid into the Syringe. Remove air bubbles as required.
Complete filling by pulling down on the plunger until red O-ring is below the ridge (fill line).
Snap off end of plunger at neck and discard.
3. Load the Injector
Hold Injector as shown.
Remember to keep thumb on Multi-function Button.
While pressing down on the Multi-function Button, insert Syringe assembly, red O-ring first, into Injector until you hear a “click” locking it into place.
Keeping thumb on Multi-function Button, invert assembly and remove Syringe from Vial/Filling Adapter with a counterclockwise turn.
4. Give Injection
Thumb off Multi-function Button
Intramuscular Injection OR
Place Syringe flat against the injection site at a 90 degree angle. Apply gentle pressure until blue retractable core stops moving.
Press down on the Multi-function Button to inject. A “click” will be heard upon injection.
Subcutaneous Injection OR
Pinch up skin over the injection site to limit depth of penetration.
Press down on the Multi-function Button to inject. A “click” will be heard upon injection.
Immediately following injection apply pressure to injection site for approximately 60 seconds.
Hold Multi-function Button and recap Syringe.
Push Button inward to eject Syringe.
Discard used Syringe.